Compact x-ray microtomography system for element mapping and absorption imaging

2007 
We have designed and built a compact x-ray microtomography system to perform element mapping and absorption imaging by exploiting scanning fluorescence tomography and full-field transmission microtomography, respectively. It is based on a low power microfocus tube and is potentially appropriate for x-ray diagnostics in space. Full-field transmission tomography yields the three-dimensional inner structure of an object. Fluorescence microtomography provides the element distribution on a virtual section through the sample. Both techniques can be combined for appropriate samples. Microradiography as well as fluorescence mapping are also possible. For fluorescence microtomography a small and intensive microbeam is required. It is generated using a polycapillary optic. Operating the microfocus tube with a molybdenum target at 12W, a microbeam with a full width at half maximum lateral extension of 16μm and a flux of about 108photons∕s is generated. As an example of application, this beam is used to determine the...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    29
    References
    9
    Citations
    NaN
    KQI
    []