Evaluation of Small Mass Spectrometer Systems

2002 
This work is aimed at understanding the aspects of designing a miniature mass spectrometer (MS) system. A multitude of commercial and government sectors, such as the military, environmental agencies and industrial manufacturers of semiconductors, refrigerants, and petroleum products, would find a small, portable, rugged and reliable MS system beneficial. Several types of small MS systems are evaluated and discussed, including linear quadrupole, quadrupole ion trap, time of flight and sector. The performance of each system in terms of accuracy, precision, limits of detection, response time, recovery time, scan rate, volume and weight is assessed. A performance scale is setup to rank each systems and an overall performance score is given to each system. All experiments involved the analysis of hydrogen, helium, oxygen and argon in a nitrogen background with the concentrations of the components of interest ranging from 0-5000 part-per-million (ppm). The relative accuracies of the systems vary from < 1% to approx. 40% with an average below 10%. Relative precisions varied from 1% to 20%, with an average below 5%. The detection limits had a large distribution, ranging from 0.2 to 170 ppm. The systems had a diverse response time ranging from 4 s to 210 s as did the recovery time with a 6 s to 210 s distribution. Most instruments had scan times near, 1 s, however one instrument exceeded 13 s. System weights varied from 9 to 52 kg and sizes from 15 x 10(exp 3)cu cm to 110 x 10(exp 3) cu cm.