language-icon Old Web
English
Sign In

Miniature mass spectrometer

A miniature mass spectrometer (MMS) is a type of mass spectrometer (MS) which has small size and weight and can be understood as a portable or handheld device. Current lab-scale mass spectrometers however, usually weigh hundreds of pounds and can cost on the range from thousands to millions of dollars. One purpose of producing MMS is for in situ analysis. This in situ analysis can lead to much simpler mass spectrometer operation such that non-technical personnel like physicians at the bedside, firefighters in a burning factory, food safety inspectors in a warehouse, or airport security at airport checkpoints, etc. can analyze samples themselves saving the time, effort, and cost of having the sample run by a trained MS technician offsite. Although, reducing the size of MS can lead to a poorer performance of the instrument versus current analytical laboratory standards, MMS is designed to maintain sufficient resolutions, detection limits, accuracy, and especially the capability of automatic operation. These features are necessary for the specific in-situ applications of MMS mentioned above. A miniature mass spectrometer (MMS) is a type of mass spectrometer (MS) which has small size and weight and can be understood as a portable or handheld device. Current lab-scale mass spectrometers however, usually weigh hundreds of pounds and can cost on the range from thousands to millions of dollars. One purpose of producing MMS is for in situ analysis. This in situ analysis can lead to much simpler mass spectrometer operation such that non-technical personnel like physicians at the bedside, firefighters in a burning factory, food safety inspectors in a warehouse, or airport security at airport checkpoints, etc. can analyze samples themselves saving the time, effort, and cost of having the sample run by a trained MS technician offsite. Although, reducing the size of MS can lead to a poorer performance of the instrument versus current analytical laboratory standards, MMS is designed to maintain sufficient resolutions, detection limits, accuracy, and especially the capability of automatic operation. These features are necessary for the specific in-situ applications of MMS mentioned above. In typical mass spectrometry, MS is coupled with separation tools like gas chromatography, liquid chromatography or electrophoresis to reduce the effect of the matrix or background and improve the selectivity especially when the analytes are widely different in concentration. Sample preparation including sample collection, extraction, pre-separation increases the size of the mass analysis system and adds time and sophistication to the analysis. A lot of contribution promotes miniaturizing devices and simplifying the operations. A micro-GC has been implemented to fit to a portable MS system. Besides microfluidics is a competent candidate for MMS and automating sample preparation. In this technique, most of the steps for sample preparation are staged similarly with laboratory systems, but miniature chip-based devices are used with low consumption of sample and solvents. One way to circumvent classical, lab-based sample introduction systems is the use of ambient ionization as it does not require mechanical or electrical coupling to a MMS and can generate ions in the open atmosphere without prior sample preparation. Different ambient ionization methods, including low-temperature plasma, paper spray, and extraction spray, have been demonstrated to be highly compatible with MMS.Without separation coupling, the basic building blocks in MMS, which are similar in composition with the conventional laboratory counterpart, are sample inlet, ionization source, mass analyzers, detector, vacuum system, instrument control and data acquisition system.Three most important components in MMS contributing to miniaturization are mass analyzer, vacuum system and electronics control system. Reducing the size of any components is beneficial to the miniaturization. However, it is noticeable that minimizing the analyzer’s size can greatly enhance the miniaturization of the other components especially the vacuum system because the analyzer is the pressure deciding factor for MS analysis and pressure interface fabrication. The smaller the mass analyzer requires smaller control system to generate adequate electric field and magnetic field strength, which are two fundamental fields separating ions based on their mass-to-charge ratio. Because a compact circuit can generate a high electric field, decreasing the size of the voltage-generating system does not significantly affect to the miniaturization of time-of-flight mass spectrometry (TOF) and electric sectors which use only the electric field to separate ions. In principle, the electromagnetic field mainly depends on the shape of the mass analyzers. As a result, a smaller magnet fitting with small size MS reduces the system weight significantly. In practice, when reducing the size, the geometries of mass analyzer are distorted. For example, smaller volume in ion trap leads to lower trapping capacity and therefore results in a loss of resolution and sensitivity. However, by operating tandem MS which is capable to give the structural information of the sample from a complex mixture, resolution and selectivity are greatly enhanced. In general, beam-type mass analyzers, such as TOF and sector mass analyzers, are much larger than ion trap type such as Paul trap, Penning trap or Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR). Additionally, ion trap mass analyzers can be used to perform multistage MS/MS in a single device. As a result, ion traps are received dominant attention for building a MMS. Some researchers are successful in designing a series of miniature TOF mass analyzers. Cotter at Johns Hopkins University used a pulsed extraction in linear time of flight mass analyzer and the ions are accelerated to with higher energy of 12 keV to enable detection of high-mass. The group achieved resolutions of 1/1200 and 1/600 at m/z 4500 and 12000 respectively. This mini analyzer can measure 66k Da proteins, mixtures of oligonucleotides, and biological spores. Verbeck at University of North Texas, created a mini-TOF based on reflectron TOF with a microelectromechanical system technology. To overcome the low resolution of short flight tube, the effective ion travelling path length is extended by moving ions back and forth in periods of time. The system used a 5-cm endcap reflectron TOF with higher-order kinetic energy focusing to analyze the ions with m/z exceeding 60,000. Ecelberger, a senior professional staff scientist in the Sensor Science Group of the Research and Technology Development Center at APL also developed a suitcase TOF incorporated with matrix-assisted laser desorption/ionization MALDI. The suitcase TOF was tested by scientists from U.S. Army Soldier and Biological Chemical Command. The samples are biological toxins and chemical agents with the mass range from a few hundred daltons to over 60 kDa. The Suitcase TOF was referenced with a commercial TOFMS for the same experiments. Both instruments can detect all but a few compounds with very encouraging results. Because a commercial TOFMS uses a higher voltage pulsed extraction with longer flight tube with other optimized conditions, it generally has better sensitivity and resolution than a suitcase TOF. However, in the case of very high mass compounds, the suitcase TOF shows as good resolution and sensitivity as the commercial TOF. The suitcase TOF was also tested with a series of chemical weapons agents. Every compound tested was detected at levels comparable to standard analytical techniques for these agents. Several miniature double-focusing mass analyzers have been fabricated. A nonscanning Mattauch–Herzog geometry sector was developed using new materials to construct a lighter magnet. Under the collaboration of University of Minnesota and Universidad de Costa Rica, a miniature double-focusing sector was produced under sophisticated technique of conventional machining methods and thin film patterning to overcome the distortion of the electric-magnetic fields due to small size. The MMS can reach a detection limit close to 10 ppm, a dynamic range of 5 orders of magnitude and a mass range up to 103 Da. The mass analyzer overall sizes 3.5cmx6cmx7.5 cm and it weighs 0.8 kg and consumes 2.5 W. The linear quadrupole mass filter or quadrupole mass analyzer is one of the most popular mass analyzer. The mini-quadrupole has been used as a single analyzer or in arrays of identical mass analyzers. The quadrupole array has rods of 0.5mm radius and 10mm long while another one with rods of 1mm radius and 25mm long. These mini quadrupole were developed and characterized at a radio frequency (RF) higher than 11 MHz. Volatile organic compounds were ionized by electron ionization and were characterized with unit resolution. Micromachining was applied to produce a much smaller V-groove quadrupole. Ion traps include quadrupole ion traps or Paul trap, Fourier transform ion cyclotron resonance or Penning trap and newly developed orbitrap. However, Paul trap receives a great focus from researchers for a MMS because of its distinct advantages over other mass analyzers for building MMS. One of the benefits is that ion traps can work at much higher pressures than beam type mass analyzers and can be simplified with different geometry for the ease of fabrication. For example, a miniature quadrupole ion trap mass analyzers, such as cylindrical ion trap, linear ion trap, rectilinear ion trap), can operate at several mTorr in contrast to 10−5 Torr or less for other analyzers and it is able to perform MS/MS in a single device with minimum size of electronics system. Nevertheless, as the size gets smaller, it is hard to maintain the electric field shape and precise configuration and will negatively affect ion motion. The goal is to make the trap smaller without losing ion capacity. Tridion-9 mass spectrometer with toroidal ion trap is designed with a doughnut-shaped volume that can hold up to 400 times more ions. The outstanding result is achieved as the radius is reduced to one-fifth of a conventional laboratory ion trap while maintaining the ion capacity.

[ "Ion trap", "Ion source", "Ionization" ]
Parent Topic
Child Topic
    No Parent Topic