Efficacy of the RemoweLL cardiotomy reservoir for fat and leucocyte removal from shed mediastinal blood: a randomized controlled trial.
2016
Introduction:Re-transfusion of lipid particles and activated leucocytes with shed mediastinal blood (SMB) can aggravate cardiopulmonary bypass-associated inflammation and increase the embolic load. This study evaluated the fat and leucocyte removal capacity of the RemoweLL cardiotomy reservoir.Methods:Forty-five patients undergoing elective on-pump cardiac surgery were randomly allocated to filtration of SMB using the RemoweLL or the Admiral cardiotomy reservoir. The primary outcome was a drop in leucocytes and lipid particles obtained with the two filters. The effect of the filters on other blood cells and inflammatory mediators, such as myeloperoxidase (MPO), was also assessed.Results:The RemoweLL cardiotomy filter removed 16.5% of the leucocytes (p<0.001) while no significant removal of leucocytes was observed with the Admiral (p=0.48). The percentage reductions in lipid particles were similar in the two groups (26% vs 23%, p=0.2). Both filters similarly affected the level of MPO (p=0.71).Conclusion:Th...