A novel microactuator for microbiopsy in capsular endoscopes
2008
This paper presents a LiGA (a German acronym for lithographie, galvanoformung, abformung) process based microactuator to be used for microbiopsy in capsular endoscopes. This microactuator is designed to be integrated into a capsular endoscope and to extract tissue samples inside the small gastrointestine which a conventional endoscope cannot reach. The proposed microactuator was fabricated as a cylindrical shape of diameter 10 mm and length 1.8 mm. This actuator consists of three parts: a microbiopsy part with a microspike, an actuating part with a torsion spring and a triggering part with a shape memory alloy (SMA) heating wire and polymer string. In order to extract sample tissue, a microspike in the developed actuator moves forward and backward using the slider-crank mechanism. For low power consumption triggering, a polymer-melting scheme was applied. The SMA heating wire consumed approximately 1.5 V × 160 mA × 1 second (66.67 µWH) for triggering. The precise components of the microactuator were fabricated using the LiGA process in order to overcome the limitations in accuracy of conventional precision machining. The developed microactuator was evaluated by extracting tissue samples from the small intestine of a pig ex vivo, and examining the tissue with hematoxylin and eosin (H&E) staining protocol. The experimental tests demonstrated that the developed microactuator with microspike successfully extracted tissue samples from the pig's small intestines.