Komplikationen faltbarer Intraokularlinsen mit der Folge der Explantation: Untersuchungsergebnisse der Jahre 2000 und 2001

2003 
Purpose. A survey was sent to all members of the German Society of Ophthalmic Surgeons to evaluate complications that required explantation of foldable intraocular lenses (10Ls). Methods. Data on preoperative visual acuity, foldable IOL design and material, and the reason for explantation were obtained and analysed.We received 185 surveys for 2000 and 2001. Results. In 2000 and 2001,the most common reasons for IOL explantation were: incorrect lens power(31%) and dislocation (19%) in three-piece acrylic 10Ls,incorrect power (31%) and decentration (17%) in in monofocal silicone IOLs, and opacification (46%) in hydrophilic IOLs.The most common reasons for the exchange of monofocal lOLs were lens opacification in 2000 and incorrect IOL power in 2001. Conclusion. Besides the most frequent complications of decentration and incorrect power observed in rigid IOLs, new complications associated with foldable IOis occurred (e.g., optic opacification and calcification). Accurate calculation of IOL power and further improvements in material and design are still necessary to minimize the rate of explantations.