Performance evaluation of Honeywell silicon piezoresistive pressure transducers for oceanographic and limnological measurements

2005 
Abstract Simultaneous calibrations of three temperature-compensated piezoresistive ruggedized precision “absolute” pressure transducers (Honeywell model PPTR0040AP5VB-BD), which have been designed specially for long-term coastal oceanographic and limnological measurements, have been carried out at four differing temperatures (10°, 20°, 30°, and 40°C) to evaluate their suitability for such applications. The full-scale pressure range of these shallow water absolute pressure sensors is ≈ 2800 hPa (equivalent to water depth of ≈ 18 m). Measurement results have been used to examine the transducers’ performance indicators, such as zero-point offset, accuracy, linearity, hysteresis, temperature sensitivity, and slope. Differing piezoresistive ruggedized precision absolute pressure transducers (PPTRs) exhibited differing zero-point offset values, ranging from 2 to −79 hPa. Temperature sensitivity of zero-point offset was ≈0.3 hPa over the temperature range 10°–40°C. The mean hysteresis over the full-scale absolut...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    3
    References
    12
    Citations
    NaN
    KQI
    []