Towards a stable and absolute atmospheric carbon dioxide instrument using spectroscopic null method
2013
Abstract. We present a novel spectral method to measure atmospheric carbon dioxide (CO 2 ) with high precision and stability without resorting to calibration tanks during long-term operation. This spectral null method improves precision by reducing spectral proportional noise associated with laser emission instabilities. We employ sealed quartz cells with known CO 2 column densities to serve as the permanent internal references in the null method, which improve the instrument's stability and accuracy. A prototype instrument – AB solute C arbon dioxide (ABC) is developed using this new approach. The instrument has a one-second precision of 0.02 ppm, which averages down to 0.007 ppm within one minute. Long-term stability of within 0.1 ppm is achieved without any calibrations for over a one-month period. These results have the potential for eliminating the need for calibration cylinders for high accuracy field measurements of carbon dioxide.