Effect of photon counting shot noise on total internal reflection microscopy.

2021 
Total internal reflection microscopy (TIRM) measures changes in the distance between a colloidal particle and a transparent substrate by measuring the scattering intensity of the particle illuminated by an evanescent wave. From the distribution of the recorded separation distances, the height-dependent effective potential φ(z) between the colloidal particle and the substrate can be measured. In this work, we show that spatial resolution with which TIRM can measure φ(z) is limited by the photon counting statistics of the scattered laser light. We develop a model to evaluate the effect of photon counting statistics on different potential profiles using Brownian dynamics simulations and experiments. Our results show that the effect of photon counting statistics depends on spatial gradients ∂φ/∂z of the potential, with the result that sharp features tend to be significantly blurred. We further establish the critical role of photon counting statistics and the intensity integration time τ in TIRM measurements, which is a trade-off between narrowing the width of the photon counting distribution and capturing the instantaneous position of the probe particle.