Miniature Cylindrical Ion Trap Mass Spectrometer

2002 
A miniature cylindrical ion trap mass spectrometer is described, and preliminary data are presented. Functionality and performance of laboratory-scale instruments have been maintained to the extent possible in this battery-operated mass spectrometer. Capabilities include tandem mass spectrometry experiments. Custom-designed electronic components include the RF scanning and amplification system, data acquisition components, and lens power supplies, as well as a custom-software application. Direct leak and membrane introduction inlet systems are used for sample introduction. A mass/charge range of ∼250 Th with unit mass resolution has been demonstrated.