A self-contained 3He refrigerator suitable for long duration balloon experiments

1998 
Abstract We describe the design, development and test of a self-contained 3 He fridge, aimed to cool at ≲0.3 K a bolometric detection system with a hold time of two weeks. The system is robust and suitable for operation on a long-duration stratospheric balloon payload. A physical model of the fridge has been developed, which describes the measured hold time, limiting temperature and load curve of the fridge. The system has been flight tested successfully during two balloon flights of the BOOMERanG payload in summer 1997.