Location and structure of fibrous sheath formed after placing a tunneled hemodialysis catheter in a large pig model

2018 
Background and objectives:We evaluated the location and structure of the fibrous sheath formed after the placement of tunneled, cuffed hemodialysis catheters in large animals, 70 kg pigs. We focused on describing the location of the fibrous sheath in relation to the catheter. Its location explains the fibrous sheath’s ability to cause catheter dysfunction by covering the catheter exit ports located at the catheter’s tip.Design:We used three animals. Each animal had a tunneled, cuffed, 15-French diameter hemodialysis catheter placed in the external jugular vein, with the tip at the junction of the superior vena cava and the right atrium. Two animals were sacrificed at 5 weeks and one animal at 17 weeks after catheter placement. The catheter and surrounding tissues were removed in one block. The fibrous sheath was dissected and longitudinally cut along the catheter to evaluate its extension in relation to the catheter. Relevant portions of the fibrous sheath were sent for pathology examination.Results:The f...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    15
    References
    2
    Citations
    NaN
    KQI
    []