[Ahmed glaucoma valve via pars plana access. Long-term results of implantation for therapy refractive glaucoma].

2013 
BACKGROUND: Aim of this study is the presentation of long-term results regarding the effectiveness and safety of the implantation of Ahmed glaucoma valve (AGV) devices using a pars plana approach after vitrectomy in the treatment of therapy refractive glaucoma. METHODS: The implantation of AGV devices using a pars plana approach after vitrectomy was performed in 27 eyes of 22 patients with neovascular glaucoma (n = 7 patients), uveitis (n = 6), complex juvenile secondary glaucoma (n = 4), primary open angle glaucoma (n = 3), pseudoexfoliation glaucoma (n = 1) and nanophthalmus (n = 1). Intraocular pressure, visual acuity and subsequent complications were documented. The mean follow-up period was 23.6 months. RESULTS: Intraocular pressure was significantly reduced from 30.2 mmHg preoperatively to 13 mmHg after a follow-up of 36 months. The mean visual acuity did not change significantly, five eyes required a subsequent explantation of which three were due to AGV-related complications. CONCLUSIONS: The implantation of AGV devices using a pars plana approach permits an effective and permanent reduction of intraocular pressure (IOP) even in severe therapy refractive secondary glaucoma; however, the risk of serious sight-threatening complications has to be taken into account.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    1
    Citations
    NaN
    KQI
    []