Higher order aberration comparison between two aspherical intraocular lenses: MC6125AS and Akreos advanced optics

2015 
AIM: To compare higher order aberrations in two aspherical intraocular lenses(IOLs): Akreos advanced optics(AO) and Dr. Schmidt Microcrystalline 6125 aspheric anterior surface(MC6125AS) with each other. METHODS: Forty eyes of 39 patients underwent phacoemulsification and Akreos AO and MC6125 AS were implanted in their eyes in a random manner. Three months post-operatively, higher order aberrations including spherical aberration, coma aberration, and total aberrations were measured and compared.RESULTS: The total aberration was 0.24±0.17 in eyes with Dr. Schmidt and 0.20 ±0.01 in eyes with Akreos AO(P =0.361). The mean of coma aberration was 0.17 ±0.21 and 0.09 ±0.86 in Dr. Schmidt and Akreos lenses,respectively(P =0.825). Total spherical aberration was almost the same in both groups(mean: 0.05, P =0.933).Best corrected visual acuity in Akreos AO(0.10±0.68) and Dr. Schmidt(0.09±0.67) did not differ significantly(P =0.700). CONCLUSION: There is no statistically significant difference in the higher order aberrations between these two aspherical lenses.