Reliability of the electronic early treatment diabetic retinopathy study testing protocol in children 7 to <13 years old

2003 
Abstract Purpose To assess the test-retest reliability of the electronic Early Treatment Diabetic Retinopathy Study (E-ETDRS) visual acuity algorithm using the computerized Electronic Visual Acuity (EVA) tester in children 7 to Design Test-retest reliability study. Methods This multicenter study involved 245 subjects at four clinical sites. As the main outcome measure, visual acuity was measured twice using the E-ETDRS testing protocol on the EVA system, which uses a programmed handheld device to communicate with a personal computer and a 17-inch monitor at a 3-m test distance. Results Test–retest reliability was high ( r = .94 for right eyes and 0.96 for left eyes) and for both right and left eyes, 89% of retest scores were within 0.1 logarithm of the minimal angle of resolution (logMAR) (five letters) of the initial test score and 99% of retests were within 0.2 logMAR (10 letters). Reliability was high across the age range of 7 to Conclusions The E-ETDRS protocol using the EVA has high test–retest reliability in children 7 to