Development of an atomic-force-microscope-based hanging-fiber rheometer for interfacial microrheology

2009 
A new interfacial microrheology technique using atomic force microscope AFM as a force sensor is developed. The probe used for microrheology contains a long vertical glass fiber with one end glued onto a rectangular shaped cantilever beam and the other end immersed through a water-air interface. The motion of the modified cantilever can be accurately described by the Langevin equation for a damped harmonic oscillator, from which we obtain the friction coefficient of the glass fiber in contact with the water. It is found that contains two contributions. One is generated by the bulk fluid, which increases with the immersion length of the glass fiber. The other contribution comes from the contact line between the water-air interface and the glass fiber, which is obtained by an extrapolation of the measured at the limit of zero immersion length. The experiment thus demonstrates an application of AFM in the studies of interfacial microrheology and contact line dynamics.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    39
    References
    24
    Citations
    NaN
    KQI
    []