Pantak Therapax SXT 150 : performance assessment and dose determination using IAEA TRS-398 protocol
2005
The performance assessment and beam characteristics of the Therapax SXT 150 unit, which encompass both low and medium-energy beams, were evaluated. Dose determination was carried out by implementing the International Atomic Energy Agency (IAEA) TRS-398 protocol and measuring all the dosimetric parameters in order to have a solid, consistent and reliable data set for the unit. Mechanical movements, interlocks and applicator characteristics agreed with specifications. The timer exhibited good accuracy and linearity. The output was very stable, with good repeatability, long-term reproducibility and no dependence on tube head orientation. The measured dosimetric parameters included beam first and second half-value layers (HVLs), absorbed dose rate to water under reference conditions, central axis depth dose distributions, output factors and beam profiles. Measured first HVLs agreed with comparable published data, but the homogeneity coefficients were low in comparison with typical values found in the literatu...