Atmospheric pressure ionization in a miniature mass spectrometer.

2005 
A miniature cylindrical ion trap mass spectrometer featuring an atmospheric pressure interface allowing atmospheric pressure chemical ionization and electrospray ionization is described together with its analytical performance characteristics. The vacuum system, ion optics, mass analyzer, control electronics system, and detection system have all been designed and built in-house. The design is based upon a three-stage, differentially pumped vacuum system with the instrument capable of being interfaced to many types of atmospheric pressure ionization sources. Ions are transferred through home-built ion optics, and instrument control is achieved through custom-designed electronics and LabView control software. Corona discharge ionization and electrospray ionization sources are implemented and used to allow the analysis of both gaseous- and solution-phase samples during the characterization of the instrument. An upper mass/charge limit of ∼450 Th with unit resolution was achieved using a 2.5-mm-internal radiu...