Disposable biosensor based on cathodic electrochemiluminescence of tris(2,2-bipyridine)ruthenium(II) for uric acid determination.

2013 
Abstract A new method for uric acid (UA) determination based on the quenching of the cathodic ECL of the tris(2,2-bipyridine)ruthenium(II)–uricase system is described. The biosensor is based on a double-layer design containing first tris(2,2-bipyridine)ruthenium(II) (Ru(bpy) 3 2+ ) electrochemically immobilized on graphite screen-printed cells and uricase in chitosan as a second layer. The uric acid biosensing is based on the ECL quenching produced by uric acid over the cathodic ECL caused by immobilized Ru(bpy) 3 2+ in the presence of uricase. The use of a −1.1 V pulse for 1 s with a dwelling time of 10 s makes it possible to estimate the initial enzymatic rate, which is used as the analytical signal. The Stern–Volmer type calibration function shows a dynamic range from 1.0 × 10 −5 to 1.0 × 10 −3  M with a limit of detection of 3.1 × 10 −6  M and an accuracy of 13.6% (1.0 × 10 −4  M, n  = 5) as relative standard deviation. Satisfactory results were obtained for urine samples, creating an affordable alternative for uric acid determination.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    37
    References
    31
    Citations
    NaN
    KQI
    []