Validation of an Affordable Handheld Wavefront Autorefractor

2019 
SIGNIFICANCE: There is a critical need for tools that increase the accessibility of eye care to address the most common cause of vision impairment: uncorrected refractive errors. This work assesses the performance of an affordable autorefractor, which could help reduce the burden of this health care problem in low-resource communities. PURPOSE: The purpose of this study was to validate the commercial version of a portable wavefront autorefractor for measuring refractive errors. METHODS: Refraction was performed without cycloplegia using (1) a standard clinical procedure consisting of an objective measurement with a desktop autorefractor followed by subjective refraction (SR) and (2) with the handheld autorefractor. Agreement between both methods was evaluated using Bland-Altman analysis and by comparing the visual acuity (VA) with trial frames set to the resulting measurements. RESULTS: The study was conducted on 54 patients (33.9 ± 14.1 years of age) with a spherical equivalent (M) refraction determined by SR ranging from -7.25 to 4.25 D (mean ± SD, -0.93 ± 1.95 D). Mean differences between the portable autorefractor and SR were 0.09 ± 0.39, -0.06 ± 0.13, and 0.02 ± 0.12 D for M, J0, and J45, respectively. The device agreed within 0.5 D of SR in 87% of the eyes for spherical equivalent power. The average VAs achieved from trial lenses set to the wavefront autorefractor and SR results were 0.02 ± 0.015 and 0.015 ± 0.042 logMAR units, respectively. Visual acuity resulting from correction based on the device was the same as or better than that achieved by SR in 87% of the eyes. CONCLUSIONS: This study found excellent agreement between the measurements obtained with the portable autorefractor and the prescriptions based on SR and only small differences between the VA achieved by either method.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    24
    References
    7
    Citations
    NaN
    KQI
    []