Determination of triethanolamine in air samples by gas chromatography-mass spectrometry

1998 
Low amounts of triethanolamine, collected in ORBO 53 tubes during air sampling, required the development of a very sensitive method for determination. After desorption and silylation reaction with trimethylsilyl imidazole/trimethyl chlorosilane, the derivative was analyzed by gas chromatography-mass spectrometry on an Ultra 2 silica capillary column in single ion monitoring mode (retention time: about 6 min). The method has a detection limit of 1–2 pg with a desorption efficiency of about 81%. Linearity of response was ascertained in the ranges 10–100 ng and 100–1000 ng. Short-term method validation was carried out by intra- and inter-day assays on three amounts for each reference calibration curve. All results satisfied the pre-defined acceptance criteria. In general, the whole procedure was easily performed and was appropriate for our needs. Breakthrough volume was appropriate for our needs. Breakthrough volume was determined on authentic samples and was about 40–60 L, using a flow rate of 1 L·min−1. The amounts of triethanolamine found in the samples ranged from 150 to 250 ng (about 2.5–4.2 μg·m−3).