Blood-compatible biomaterials by surface coating with a novel antithrombin–heparin covalent complex

2002 
Abstract Covalent antithrombin–heparin complex (ATH) was covalently grafted to a polycarbonate urethane (Corethane ® ) endoluminal graft (a kind gift of Corvita Corporation) after being activated using 0.3% m/m NaOCl in 0.15  m phosphate pH 6.0. ATH graft density (1.98×10 −7  mol/m 2 ) was 6 times the maximum amount of unfractionated heparin (UFH) that could be bound to polycarbonate urethane surfaces. Surface-bound ATH could be stored in sterile 0.15  m NaCl at 4°C for at least 2 months with good antithrombotic activity before being implanted into rabbits. Analysis of ATH-coated tubing showed that it contained significant direct thrombin inhibitory activity. In vivo testing in a rabbit model was compared to non-activated non-coated surfaces, activated-non-coated surfaces, hirudin-coated surfaces and antithrombin (AT)-coated surfaces. The weight of the clot generated in the ATH-coated graft tubing was significantly less than the weight of the clot generated within the hirudin-coated graft ( p =0.03 with a 1-tailed Student's t test). The anticoagulant nature of ATH grafts in vivo was shown to be due to bound ATH because both the AT-coated surfaces and non-coated but activated surfaces showed similar thromboresistant efficacy to that of untreated material (ANOVA; p 125 I-AT binding sites (⩾1.5×10 −8  mol/m 2 ). Thus, ATH appears to be a good candidate for coating cardiovascular devices, such as endoluminal grafts, with high levels of substitution and significant long-term blood-compatibility.