Diagnostic timing system for the TMX-Upgrade

1981 
This system provides trigger signals at various times and clock signals at various frequencies for the CAMAC transient recorders of the plasma diagnostics system for the TMX-Upgrade. The timing system is designed so that all clocks are in fixed-phase relation to their corresponding triggers and to each other. Therefore, data recorded from the different diagnostics can be directly time compared. Trigger signals can be generated in 100-ns increments, with an uncertainty of 500 ps. The clock signals have a time uncertainty of less than 1 ns. The system is arranged so that these accuracies are maintained over the entire diagnostic room. The timing system is modular and uses mostly digital delay generators, signal fan outs, and frequency dividers. Because of the modular approach, the system can be arranged in several ways (producing many possible trigger times and sample rate clocks) and still maintain a system in which all clocks and triggers are in a fixed-phase relationship.