AngioVac suction embolectomy using a roller pump: Operative technique.

2021 
AngioVac suction embolectomy is recommended by the manufacturer to be performed with a centrifugal pump due to safety considerations. However, roller head pumps are significantly cheaper to use, and thus may be more readily available during shortages and in resource-poor settings. We present the technique of Angiovac suction embolectomy being successfully performed with a roller pump to evacuate a clot-in-transit in the inferior vena cava and right atrium, along with discussion of important safety caveats.