Rapid and sensitive quantitation of heme in hemoglobinized cells

2016 
Rapid and accurate heme quantitation in the research lab has become more desirable as the crucial role that intracellular hemoproteins play in metabolism continues to emerge. Here, the time-honored approaches of pyridine hemochromogen and fluorescence heme assays are compared with direct absorbance-based technologies using the CLARiTY spectrophotometer. All samples tested with these methods were rich in hemoglobin-associated heme, including buffered hemoglobin standards, whole blood from mice, and murine erythroleukemia (MEL) and K562 cells. While the pyridine hemochromogen assay demonstrated the greatest linear range of heme detection, all 3 methods demonstrated similar analytical sensitivities and normalized limits of quantitation of ∼1 µM. Surprisingly, the fluorescence assay was only shown to be distinct in its ability to quantitate extremely small samples. Using the CLARiTY system in combination with pyridine hemochromogen and cell count data, a common hemoglobin extinction coefficient for blood and ...