Use of a comprehensive numerical model to improve biventricular pacemaker temporization in patients affected by heart failure undergoing to CRT-D therapy
2010
Cardiac resynchronization therapy by biventricular pacemaker/ICD implantation is a validated therapy for patients affected by heart failure with asynchrony of ventricular contraction. Considering the large number of parameters which play a role in cardiac resynchronization therapy, a comprehensive numerical model of cardiocirculatory system could be a useful tool to support clinical decisions. A variable elastance model of ventricles was updated to model the interventricular septum and to simulate the interventricular and the intraventricular desynchrony, and the effect of the biventricular stimulation. In addition, a numerical model of the biventricular pacemaker, which drives the beginning of the heart chambers and interventricular septum contraction, was also developed. In order to validate the model,