A multicentre european evaluation of the Kodak EKTACHEM GLU/BUN analyzer using NCCLS guidelines and other approaches.

1982 
Summary: This paper describes the field performance of the Kodak EKTACHEM GLU/BUN Analyzer for glucose and urea. NCCLS protocols PSEP-2,3 and 4 were used which enable manufacturers to establish perfonnance Claims concerning the precision and accuracy of an analytical System. This multicentre trial used four analysers in four European countries, United Kingdom, France, West Germany and Italy to assess within and between laboratory performances. Freeze dried control materials were used for the performance check experimerit PSEP-2 and for the replication experiment PSEP-3. The replication study, although very time consuming, was straight forward to undertake and the results were comparable between centres. Compliance with the protocol fbr comparison of methpds experinient PSEP-4 in which laboratories used their own hospital patient samples was more difficult. The problems with obtaining suitable samples and the performance of compaiative methods are discussed in detail.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    5
    References
    3
    Citations
    NaN
    KQI
    []