Diagnostic Accuracy of Four Levels of Manual Compression Applied in Supersonic Shear Wave Elastography of the Breast

2020 
Purpose To investigate the diagnostic accuracy of applying four levels of manual pressure in Shear Wave Elastography (SWE) of the breast and to assess inter-rater reliability. Materials and Methods Single-center prospective preliminary study including patients receiving ultrasound examination of breast lesions as part of routine clinical practice. SWE was performed on 60 breast masses (26 benign and 34 malignant) in 54 patients by a breast fellowship trained radiologist. Stiffness values were compared between benign and malignant masses at four levels of manual compression: none, mild, moderate, and marked. Accuracy of SWE was assessed using receiving operating characteristics analysis at each level. In 18 patients, a second radiologist repeated the SWE acquisitions to evaluate reproducibility. Reproducibility was assessed using intraclass correlation coefficient. Results Without compression, we observed no significant difference in stiffness (p > 0.99) between benign and malignant lesions, and SWE demonstrated low accuracy (area under the curve = 0.64). Stiffness was higher in malignant lesions at all levels of compression (p Conclusion This preliminary study suggests that not using compression during SWE for breast lesion characterization offers suboptimal results. On the contrary, application of compression yields high diagnostic performance with good interobserver agreement and, as such, should be included in routine clinical practice.