Noninvasive Assessment of Jugular Venous Pressure via Force-Coupled Single Crystal Ultrasound

2018 
Objective: we have developed a handheld device for noninvasive quantitative assessment of jugular venous pressure (JVP). Methods: we used a single crystal ultrasound coupled to a force-sensing load cell to measure JVP based on the force necessary to collapse the internal jugular vein (IJV) walls. We used a gelatin-based model system of the IJV to test the ability of single crystal ultrasound to identify the IJV and verified the cross-sectional position and diameter of the vessels with conventional imaging ultrasound. We also tested our prototype device on healthy human volunteers. Results: experiments on model system demonstrated that vessel diameters determined with single crystal ultrasound were in close agreement with the diameters derived from conventional 2-D ultrasound. Proof-of-concept human experiments demonstrate that single crystal ultrasound can detect the IJV in basal and collapsed states, as compared to gold-standard sonography (insert stats). Assessment of JVP in human volunteers was physiologically consistent with and sensitive to postural changes (supine JVP 6.6 ± 2.4 mmHg; standing JVP 4.2 ± 1.9 mmHg ( p Conclusion: noninvasive assessment of JVP could prove valuable in informing rapid clinical decision-making across various pathologies and conditions leading to derangements in intravascular volume status.