Digital Multiparameter System for Characterizing the Neutron-Gamma Field in the LR-O Experimental Reactor
2007
The present paper points toward the possibility of reducing the
proportion of analog processing in an apparatus employed to
measure the neutron and photon energy spectra. The digitization
of the detector output pulses makes it possible to process the
information contained in the pulses either instantaneously or
subsequently. The present paper deals with the method of
digitizing the leading edge of the pulses generated by a
stilbene organic scintillator or an NE-213 scintillator. The
neutron versus gamma discrimination can be carried out on the
basis of the pulse leading edge duration. Coping with this task
requires a digitizer featuring quite a high sampling frequency,
which must exceed 1 GHz. The requirements for the amplitude
digitization frequency are less severe, however, the amplitude
resolution must be better than 8 bits. Both of the above
requirements can be met with present-day components. The
digitization is expected to result in reducing the apparatus
size and improving its functionality.