Prosthesis Oversizing in Balloon-Expandable Transcatheter Aortic Valve Implantation Is Associated With Contained Rupture of the Aortic Root

2012 
Background—To retrospectively investigate the potential cause of contained rupture of the aortic root in balloon-expandable transcatheter aortic valve implantation (TAVI) by means of pre- and postinterventional multislice computed tomography. Methods and Results—Seventy-two patients (mean age 82±7 years, mean aortic valve area 0.69±0.19 cm2) underwent balloon-expandable TAVI using the EdwardsSAPIEN Transcatheter Heart Valve (23 mm, n=19; 26 mm, n=50; 29 mm, n=3). Aortic annulus dimensions were quantified by multislice computed tomography–based cross-sectional area assessment and average diameter calculation (CAAD) before and after TAVI. Post-TAVI multislice computed tomography data sets were available in 65 patients; contained aortic root rupture was diagnosed in 3 patients. Pre-TAVI CAAD was 23.1±1.8 mm; post-TAVI CAAD was 22.9±1.3 mm. Median relative change in CAAD pre- and post-TAVI was −0.5% (interquartile range, 3.6%). Relative increase of 5% to 10% was observed in 4 patients (1 with contained ruptur...