Low-dose dental computed tomography: significant dose reduction without loss of image quality.

2004 
Purpose: To measure and reduce the patient dose during computed tomography (CT) for dental applications. Material and Methods: Lithium fluoride thermoluminescent dosimeters were implanted in a tissue-equivalent humanoid phantom (Alderson-Rando-Phantom) to determine doses to the thyroid gland, the active bone marrow, the salivary glands, and the eye lens. Dental CT was performed with spiral CT and a dental software package. The usual dental CT technique was compared with a new dose-reduced protocol, which delivered best image quality at lowest possible radiation dose, as tested in a preceding study. Image quality was analysed using a human anatomic head preparation. In addition, the radiation dose was compared with panoramic radiography and digital volume tomography (DVT). Eight radiologists evaluated all images in a blinded fashion. A Wilcoxon rank pair test was used for statistical evaluation. Results: Radiation dose could be reduced by a factor of 9 (max.) with the new dose-reduced protocol (e.g. bone m...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    19
    References
    55
    Citations
    NaN
    KQI
    []