Cardiac resynchronization therapy: a comparison among left ventricular bipolar, quadripolar and active fixation leads
2018
We evaluated the performance of 3 different left ventricular leads (LV) for resynchronization therapy: bipolar (BL), quadripolar (QL) and active fixation leads (AFL). We enrolled 290 consecutive CRTD candidates implanted with BL (n = 136) or QL (n = 97) or AFL (n = 57). Over a minimum 10 months follow-up, we assessed: (a) composite technical endpoint (TE) (phrenic nerve stimulation at 8 V@0.4 ms, safety margin between myocardial and phrenic threshold 15%). Baseline characteristics of the 3 groups were similar. At follow-up the incidence of TE was 36.3%, 14.3% and 19.9% in BL, AFL and QL, respectively (p < 0.01). Moreover, the incidence of RR was 56%, 64% and 68% in BL, AFL and QL respectively (p = 0.02). There were no significant differences in CE (p = 0.380). On a multivariable analysis, “non-BL leads” was the single predictor of an improved clinical outcome. QL and AFL are superior to conventional BL by enhancing pacing of the target site: AFL through prevention of lead dislodgement while QL through improved management of phrenic nerve stimulation.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
22
References
297
Citations
NaN
KQI