Design of Myoelectric Control Command of Electric Wheelchair as Personal Mobility for Disabled Person

2019 
Paralysis is a condition that causes the sufferer to be unable to move the leg muscles or hand muscles. However, a paralyzed subject can still use the electrical activity of other intact muscles as a control command for the electric wheelchair. This system includes instrumentation and processing of electromyography (EMG) signal, and interpretation for controlling the wheelchair to move forward, backward, stop, turn left, and turn right. The EMG signal of the Flexor Carpi Radialis muscles was measured and amplified by a designed instrumentation amplifier. Then, the signal was filtered to eliminate noise and maintain the EMG frequency range of 20500 Hz. After filtering, the signal was processed more to produce a linear envelope and hold the maximum value of the envelope of the EMG signal. Interpretation of command from subject was determined by thresholing method to the envelope of the EMG signal. The results of wheelchair navigation with basic movements had an average success rate of 100% of 8 subjects. The test results past the determined track have an average success rate of 95.83% for 3 subjects. It is expected that in the future, the subjects can use the wheelchair with subject’s intention speed.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    3
    References
    4
    Citations
    NaN
    KQI
    []