Determination of Arylamine Compounds in Waste Water Using Solid-Phase Extraction and Reversed-Phase High Performance Liquid Chromatography

1998 
Abstract A simple, rapid, and sensitive method for determining 10 arylamine compounds in wastewater sample using solid-phase extraction (SPE) and high performance liquid chromatography (HPLC) with UV detection is developed. An SPE procedure is described for extracting and preconcentrating the arylamines on Sep-Pak C18 cartridges with methanol as solvent. A mobile phase gradient of acetonitrile-water was used for the separation and determination of the arylamines. The detection limits of the compounds were in the range of 0.1−1.5 μg/L. Recoveries were 84−107% for 10 arylamines at concentration levels of 20 μg/L with relative standard deviations of less than 8% (n = 4).
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    8
    References
    2
    Citations
    NaN