Comparison of Ion-Pairing and Reversed Phase Liquid Chromatography in Determination of Sulfamethoxazole and Trimethoprim

2008 
Abstract Two simple, rapid, and sensitive HPLC methods have been developed for the simultaneous determination of sulfamethoxazole and trimethoprim in their pure and dosage forms, one utilizing reversed phase HPLC and the other ion-pair HPLC. In the reversed phase HPLC method (A) the mobile phase consists of 0.05% aqueous solution of formic acid with pH adjusted to 4.5±0.2 with triethylamine : acetonitrile:tetrahydrofuran 50 : 49 : 1 (v/v), and the mobile phase pumped at flow rate of 1.0 ml min−1. An Appolo LC18 column (5.0 µm), 250 mm length × 4.6 mm diameter, was utilized as the stationary phase. Detection was affected spectrophotometrically at 254 nm. In the ion-pair HPLC method (B) the mobile phase consisted of methanol : buffer 35 : 65 (v/v) with the buffer composed of potassium dihydrogen phosphate 0.3 M and sodium heptan sulfonic acid 5.0 mM. To 500 ml of buffer was added 2.0 ml triethylamine, and then the pH was adjusted to 5.0 with phosphoric acid, and the mobile phase was pumped at a flow rate of...