Molar activity – The keystone in 11C-radiochemistry: An explorative study using the gas phase method

2018 
Abstract Introduction Radiochemists/radiopharmacists, involved in the preparation of radiopharmaceuticals are regularly confronted with the requirement of continuous high quality productions in their day-to-day business. One of these requirements is high specific or molar activity of the radiotracer in order to avoid e.g. receptor saturation and pharmacological or even toxic effects of the applied tracer for positron emission tomography. In the case of 11 C-labeled radiotracers, the reasons for low molar activity are manifold and often the search for potential 12 C-contaminations is time-consuming. Methods In this study, diverse 12 C-contaminations were analyzed and quantified, which occurred during >450 syntheses of six PET tracers using [ 11 C]CO 2 or [ 11 C]CH 3 I generated via the gas phase method in a commercially available synthesizer. Additionally, non-radioactive syntheses were performed in order to identify the origins of carbon-12. Results The manifold contributions to low molar activity can be attributed to three main categories, namely technical parameters ( e.g. quality of target gases, reagents or tubings), inter/intralaboratory parameters ( e.g. maintenance interval, burden of the module, etc. ) and interoperator parameters ( e.g. handling of the module). Conclusion Our study provides a better understanding of different factors contributing to the overall carbon load of a synthesis module, which facilitates maintenance of high molar activity of carbon-11-labeled radiopharmaceuticals.