Consistency of renal stone volume measurements across CT scanner model and reconstruction algorithm configurations
2017
OBJECTIVE. The objective of this prospective study is to evaluate the consistency of renal stone volume estimation using dual-energy CT across scanner model and reconstruction algorithm configurations. SUBJECTS AND METHODS. Patients underwent scanning with routine kidney stone composition protocols on both second- and third-generation dual-source CT scanners. Images were reconstructed using filtered back projection and iterative reconstruction (IR). In addition, a modified IR kernel on the third-generation CT scanner was evaluated. Individual kidney stone volumes were determined and compared. RESULTS. No significant difference was noted in measured volumes between filtered back-projection data, IR data from the second-generation scanner, and the modified IR kernel data (p > 0.05). The third-generation commercially available IR kernel yielded lower volumes than did the other configurations (p < 0.0001). CONCLUSION. With the use of a modified kernel for the third-generation scanner, patients being monitored...