A Comparative Assessment of the Nova Stat Profile Prime Plus® Critical Care Analyzer
2020
BACKGROUND Point-of-care testing (POCT) plays an integral role in the management of acutely ill patients presenting to the emergency department (ED). Due to its rapid turnaround time, POCT has been shown to improve ED workflow, reduce unnecessary admissions and lessen the burden on ED staff. The aim of the study was to compare the accuracy, precision and linearity of the Nova Stat Profile Prime Plus® (Nova Biomedical, Waltham, MA, USA) to the Radiometer ABL800 FLEX® (Radiometer South Africa Pty Ltd, Gauteng) and the Abbott i-stat Chem8+® (Abbott, Princeton, NJ, USA) POCT analyzers. METHODS A convenience sample of 150 discarded whole blood specimens was obtained and analyzed. Paired test measurements were conducted for method comparison. Accuracy was measured by pairing individual results from the Nova Stat Profile Prime Plus® with either the Radiometer ABL800 FLEX® or the Abbot i-stat Chem8+® analyzers by calculating the differences. RESULTS The with-in run percentage coefficient of variation (%CV) was below 2.4% for pH, carboxyhemoglobin (COHb), deoxyhemoglobin (HHb), total hemoglobin (tHb), total bilirubin (tBil), sodium (Na), potassium (K), chloride (Cl), ionized calcium (iCa), urea, glucose and lactate, and was below 5.1% for all other analytes. The day-to-day %CV was below 1.6% for pH, COHb, HHb, tHb, tBil, Na, K, Cl, iCa, urea, glucose and lactate, and below 6.10% for all other analytes. The correlation coefficient (r) was 0.351 and ranged from 0.897 to 0.998 for all analytes. The mean bias was minimal for all analytes. CONCLUSION There was a good correlation between the Nova Stat Profile Prime Plus® and the Radiometer ABL800 FLEX®/Abbott i-STAT Chem8+® POCT analyzers. The Stat Profile Prime Plus® exhibited good precision both within-run and day-to-day.
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
11
References
0
Citations
NaN
KQI