Comparison of the clinical performance of three digital mammography systems in a breast cancer screening programme

2012 
This study compares the clinical performance of three digital mammography system types in a breast cancer screening programme. 28 digital mammography systems from three different vendors were included in the study. The retrospective analysis included 238 182 screening examinations of females aged between 50 and 64 years over a 3-year period. All images were double read and assigned a result according to a 5-point rating scale to indicate the probability of cancer. Females with a positive result were recalled for further assessment imaging and biopsy if necessary. Clinical performance in terms of cancer detection rate was analysed and the results presented. No statistically significant difference was found between the three different mammography systems in a population-based screening programme, in terms of the overall cancer detection rate or in the detection of invasive cancer and ductal carcinoma in situ. This was shown in both prevalent and subsequent screening examination categories. The results demonstrate comparable cancer detection performance for the three imaging system types operational in the screening programme.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    17
    References
    11
    Citations
    NaN
    KQI
    []