Assessment of the diagnostic performances of IgA heavy and light chain pairs in patients with IgA monoclonal gammopathy.

2013 
Abstract Objectives Preliminary results of the IgA Hevylite™ assay including the establishment of the 95% reference interval and assessment of the specificity and sensitivity in different populations are reported. Design and methods The concentrations of IgA heavy and light chains (HLC) enabling to determine an IgAκ/IgAλ ratio were quantified in 119 apparently healthy individuals to generate 95% reference intervals. The specificity of this assay was assessed in 48 patients with an isolated polyclonal IgA increase. In a retrospective analysis of 68 patients with a monoclonal component type IgA (MC-IgA) identified by serum immunofixation (IFE), IgA HLC ratio values were compared with known results for serum protein electrophoresis (SPE) and free light chain (FLC) ratios. Results The 95% reference range obtained in 119 controls (0.91–2.04) was close to that quoted by the manufacturer (0.80–2.04). Eight of the 48 patients (16.7%) with a polyclonal IgA increase had an IgA HLC ratio above the upper limit of the 95% reference interval. The IgA HLC ratio identified 65 (95.6%) among 68 patients with MC-IgA identified on the basis of IFE. For 34 of these patients (50%), MC-IgA was not detected by SPE due to its co-migration with alpha-2 or beta-globulins. Conclusions Compared with serum IFE, the IgA HLC ratio has a sensitivity of 95.6%. Further studies are needed to assess the specificity of the IgA HLC ratio in patients with an isolated polyclonal increase of serum IgA.