Calibration and Testing of Small High-Resolution Transition Edge Sensor Microcalorimeters with Optical Photons
2021
Pulses of narrow line-width optical photons can be used to calibrate and test sub-2 eV full-width at half-maximum (FWHM) energy resolution transition-edge sensor (TES) microcalorimeters at low energies ( $ keV), where it is very challenging to obtain X-ray calibration lines comparable to (or narrower than) the detector resolution. This scheme depends on the ability to resolve the number of 3 eV photons in each pulse, which we have recently demonstrated up to photon numbers of about 300. At LTD-18 we showed preliminary results obtained with this technique on a 0.25 eV baseline resolution TES microcalorimeter designed for the ultra-high-resolution sub-array of the Lynx mission. The line-shape was well described by a simple Gaussian. However, the difficulty of delivering photons to the small 46 $\mathrm{\mu m}$ square absorbers resulted in a large thermal crosstalk signal, whose random nature is expected to rapidly degrade the observed energy resolution towards higher photon numbers/energies. We have since improved the coupling between the optical fiber and the TES absorber and report here our current results.