REDUCING THE DISTAL PROFILE OF DUAL MOBILITY LINERS CAN MITIGATE SOFT-TISSUE IMPINGEMENT AND LINER ENTRAPMENT WITHOUT AFFECTING MECHANICAL PERFORMANCE

2016 
Introduction Dual Mobility (DM) implants have gained popularity for the treatment and prevention of hip dislocation, with increased stability provided by a large diameter mobile liner. However, distal regions of the liner can impinge on soft-tissues like hip capsule and iliopsoas, leading to anterior hip pain. Additionally, soft-tissue impingement may trap the mobile liner, leading to excessive loading of the liner rim, from engagement with the femoral stem, and subsequent intra-prosthetic dislocation. The hypothesis of this study was that reducing the liner profile below the equator (contoured design) can mitigate soft-tissue impingement without compromising inner-head pull-out resistance and overall hip joint stability ( Fig. 1 ). Methods The interaction of conventional and contoured liners with anterior soft-tissues was evaluated in 10 cadaveric hips (5 specimens; 2 male, 3 female; age 65 ± 10 yrs; liner diameter 42–48mm) via visual observation and fluoroscopic imaging. A metal wire was sutured to the deep fibers of the iliopsoas tendon/muscle, and metal wires were embedded in the mobile liners for fluoroscopic visualization (Fig. 2) . All soft-tissue except the anterior hip capsule and iliopsoas was removed, and a rope was attached to the iliopsoas to apply tension along its natural orientation. Resistance to inner-head pull-out was evaluated via Finite Element Analysis (FEA) by simulating a full cycle of insertion of the inner head into the mobile liner and subsequent pullout. The femoral head, acetabular shell, and stem were modeled as rigid, while the mobile liner was modeled as plastically deformable. Hip joint stability was evaluated by dynamic simulations in for two dislocation modes: (A) Posterior dislocation (at 90° hip flexion) with internal hip rotation; (B) Posterior dislocation (starting at 90° flexion) with combined hip flexion and adduction. A 44 mm diameter conventional and a 44 mm contoured liner were evaluated during these tests. Results The cadaver experiments showed that distal portion of conventional liners impinge on anterior hip capsule and iliopsoas at low flexion angles ( (Fig. 7) . The change in tenting could be visualized as change in distance between the iliopsoas wire, and the contoured/conventional liners on sequential fluoroscopic images. The maximum reduction in iliopsoas tenting for a given specimen ranged from 1.8 mm to 5.5 mm. Additionally, the contoured and conventional liners had identical inner-head pull-out resistance (901N vs. 909N), jump distance (9.4 mm mode-A, 11.7 mm mode-B) and impingement-free range of motion (47° mode-A, 29° mode-B). Conclusion This study showed that distal portions of conventional DM liners can impinge against iliopsoas and hip capsule in low flexion leading to functional impediment of liner motion. Additionally, reducing the liner profile below the equator led to significant reduction in soft-tissue impingement/tenting without affecting mechanical performance. Thus, a contoured dual mobility liner design may reduce the risk of anterior hip pain and intra-prosthetic dislocation resulting from soft-tissue impingement and liner entrapment. To view tables/figures, please contact authors directly .
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []