Quantifying the setup uncertainty of a stereotactic murine micro-image guided radiation therapy system

2019 
Objective:Investigate the reproducibility of murine cranial positioning using solely a stereotactic stage, and quantify the potential improvements from the on-board image guidance of the X-RAD SmART irradiator.Methods:For intermouse reproducibility, athymic nude mice (N = 5, ×4 groups) were cranially fixed on a stereotactic stage. Each mouse was imaged via cone-beam CT (CBCT). A virtual isocenter target was placed in the brain, the stage shifted to that target, and the couch positions recorded. The mouse was removed from the stage and this process repeated twice (N=60 measurements). The first acquired CBCT coordinates (within each group of five mice) were used to define “stereotactic couch coordinates.” CBCT shifts were calculated to quantify the accuracy of setup based on couch coordinates alone. For intramouse reproducibility, C57BL/6 mice (N=4) were imaged daily for 7 days. Each mouse had individual stereotactic coordinates defined from their first day of CBCT localization, and positional shifts requir...