Palm Glucose Readings Compared with Fingertip Readings Under Steady and Dynamic Glycemic Conditions, Using the OneTouch® Ultra® Blood Glucose Monitoring System

2005 
Objectives: Past studies have suggested the absence of lag between palm glucose and fingertip glucose, even when glucose levels are changing rapidly. However, at any given time point, there may be differences between palm and fingertip glucose values because of glycemic instability and/or test methodology. The objectives of this study included assessing the variability in fingertip blood glucose test results between two fingers, and establishing whether the variability in blood glucose test results obtained from the palm was clinically equivalent to that observed in fingertip-to-fingertip comparisons. Methods: This multicenter trial was conducted on patients under both steady-state glycemic conditions and after meal and exercise challenges (to promote rapidly changing glucose). Sequential capillary glucose testing, performed with the One Touch® Ultra® Blood Glucose Monitoring System (LifeScan, Inc., Milpitas, CA), was allocated to two of four fingertip sites and one of two palm sites in each subject using...