Evaluation of high-pitch flash scan for pulmonary venous CTA on a 128-slice dual source CT: compared with prospective ECG-triggered sequence scan.

2013 
To compare the image quality (IQ) and radiation dose of high-pitch scan and prospective ECG-triggered sequence scan on a 128-slice DSCT system for patients with atrial fibrillation (AF). Pulmonary venous (PV) CTA was performed with two protocols, including high-pitch scan and prospective ECG-triggered sequence scan. For each protocol, 20 sex, age and body-mass-index (mean 24.2 kg/m2) matched patients were identified. Two experienced radiologists, who were blinded to the scan protocols, independently graded the CT images of the two groups by a 5-point scale for subjective IQ assessment. Measured CT attenuation (Hounsfield units ± standard deviation), signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) at various anatomic locations were also recorded for objective IQ evaluation. Radiation exposure parameters [dose length product (DLP) and effective radiation dose (ERD)] were compared. Twenty-three patients (57.5 %) showed an ECG pattern of AF in total. Subjective IQ was rated excellent in 100 % for the high-pitch scan group, while minor step artifacts were observed in two patients (10 %) with arrhythmia for the prospective ECG-triggered sequence group. There was no significant difference on IQ, neither by subjective, nor by objective measures (SNR, CNR) between the two groups. The ERD of high-pitch flash scan and prospective ECG-triggered sequence scan were 0.9 (±0.25) and 2.9 (±0.69) mSv, respectively. Significantly lower radiation was achieved by using high-pitch flash scan (P < 0.05). High-pitch flash scan can provide similar subjective and objective IQ compared with prospective ECG-triggered sequence scan for PV CTA, while radiation exposure was significantly reduced.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    23
    References
    4
    Citations
    NaN
    KQI
    []