Effect of Piezoelectric Energy on Porcine Kidneys Using the EDAP LT.02

1995 
AbstractEDAP International (Cambridge, Massachusetts) has developed a new piezoelectric lithotripter (LT.02) that differs from their earlier model (LT.01) in two important respects: method of stone visualization and available power. The LT.02 provides both in-line fluoroscopy as well as real time ultrasound. The maximum energy is 1400 Bar (compared with 1100 Bar in the LT.01). The purpose of this study was to determine whether treatment with the EDAP LT.02 piezoelectric lithotripter would cause significant renal injury in minipigs. Accordingly, 18 minipigs were divided into 3 groups of 6. Each group received a treatment of 20, 40, or 60 minutes; the power level and shock wave frequency were kept at maximal levels throughout the treatment period. Three pigs from each group were sacrificed at 72 hours (acute). The remaining 9 pigs were sacrificed 1 month following LT.02 treatment (chronic). Histopathologic analysis of the treated kidney revealed that 33 percent of the 9 acute pigs developed a small capsular...