An in vitro load evaluation of a conical implant system with 2 abutment designs and 3 different retaining-screw alloys.

2006 
Abstract The aim of this in vitro study was to evaluate the load resistance in a conical implant system by comparing combinations of 2 different abutment head angles and 3 different retaining screw materials. The retaining screw materials (titanium alloy, gold alloy, and commercially pure titanium) were tested with abutment-head angles of 20 degrees and 45 degrees. Six groups of 10 specimens each were prepared. An oblique (30-degree) compression test was performed in a Lloyd LRX universal testing machine with the abutment attached to a superstructure with a retaining screw. All specimens were loaded until fracture or permanent deformation occurred. The results were evaluated statistically with Wilcoxon signed rank test for variance distribution (P
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    37
    References
    18
    Citations
    NaN