[Quantitative determination of voglibose contents in its tablets with high-performance liquid chromatography-mass spectrometry].

2014 
OBJECTIVE: To develop a method for determination of voglibose contents in its tablets by high-performance liquid chromatography-mass spectrometry (HPLC-MS). METHODS: The measurements were carried out on an Agilent ZORBAX Eclipse Plus C18 column (2.1×150mm 3.2μm) with a temperature of 40 degrees Celsius. A mixture of methanol and water (2:3,v/v) was used as a mobile phase at a flow rate of 0.25 ml/min. Voglibose was detected in an electrospray ionization (ESI) mode with MRM. RESULTS: The calibration curves of voglibose showed good linearity in a range of 1.5804-2.6340 μg/ml (r=0.9990). The average recovery was 100.2% with RSD of 1.37% (n=6) for m/z 268.2/74.2.Linearity was obtained with r=0.9976 and the average recovery was 99.3% with RSD of 1.78% (n=6) for m/z 268.2/92.2. CONCLUSION: HPLC-MS method is accurate,reproducible and can be used for quality control of voglibose tablets.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    0
    References
    0
    Citations
    NaN
    KQI
    []