Characterization of Drug-Product-Related Impurities and Variants of a Therapeutic Monoclonal Antibody by Higher Energy C-Trap Dissociation Mass Spectrometry
2015
Mass spectrometry (MS) characterization of recombinant monoclonal antibody (mAb) drugs and their degraded and/or post-translationally modified counterparts, drug-product-related impurities and variants, is critical for successful development of biotherapeutics. Specifically in this study, drug-product-related impurities of an anti-Clostridium difficile IgG1 mAb drug substance were profiled by cation-exchange liquid chromatography (CEX) followed by the CEX peaks being fraction-collected for MS characterization. A reversed-phase liquid chromatography/mass spectrometry (LC/MS) methodology was developed on a Thermo Q-Exactive orbitrap mass spectrometer for (1) accurate mass measurements of the mAb, its CEX fractionated impurities, and their respective heavy chains and light chains and (2) middle-down LC/MS/MS of the light chains and the heavy chains using higher energy C-trap dissociation (HCD). The accurate mass measurements and the HCD middle-down MS/MS experiments identify that major impurities and variant...